3
102
Siemens-Elema AB
E382 E380E 061 01 03 01
Disassembling and assembling
Servo Ventilator 300/300A
Removing the inter-
connection cable
Patient unit connection
• Remove the pneumatic section casing as
previously described.
• Disconnect the two interconnection cable
connectors (1).
• Remove the two screws (2).
• Slide out the cable clamp (3) from the
patient unit.
Mounting the inter-
connection cable
The procedure for assembling the inter-
connection cable to the ventilator is the
reverse of the removing procedure
previously performed.
Control unit connection
• Remove the two screws (1) and lift off
the cover (2).
• Disconnect the two interconnection cable
connectors (3).
• Remove the two screws (4).
• Slide out the cable clamp (5) from the
control unit.
4
5
1
2
3
300-A48X
1
3
2
300-A47X