Servo Ventilator 300/300A
E382 E380E 061 01 03 01
Siemens-Elema AB
3
• The Servo Ventilator 300/300A must be
serviced at regular intervals by specially
trained personnel. The service intervals,
1000 and 3000 operating hours, are
described in the Operating Manual. Any
maintenance must be noted in a log
book provided for that purpose in
accordance with national regulations.
We recommend that service is done as
a part of a service contract with
Siemens.
• A 1000 hour overhaul must be performed
after 1000 hours of operation or, at the
lastest, every six months. In addition,
the ventilator shall undergo a technical
safety check (Function check) twice a
year, at six months intervals or according
to national regulations.
• A 3000 hour overhaul must be performed
after every 3000 hours of operation or,
at the latest, once every year.
• The internal batteries shall be replaced
every 3 years according to instructions in
this Service Manual. The stated battery
back-up time, approx. 30 minutes, can
be guaranteed only if they are used as
power supply back-up at mains failure.
• The battery on PC 1587
COMPUTER
INTERFACE
shall be replaced every 5 years
according to instructions in this Service
Manual.
Important
To the responsible service personnel
• The contents of this document are not
binding. If any significant difference is
found between the product and this
document, please contact Siemens for
further information.
• We reserve the right to modify products
without amending this document or
advising the user.
• Only Siemens authorized personnel shall
be permitted to service or repair the
Servo Ventilator 300/300A. Only
Siemens-Elema exchange parts or
genuine spare parts must be used. PC
boards (spare or exchange parts) must
always be kept in a package for sensitive
electronic devices.
• Siemens will not otherwise assume
responsibility for the materials used, the
work performed or any possible
consequences of same.
•
The device complies with the
requirements of the Medical
Device Directive 93/42/EEC.
• Old non-functioning batteries
and O
2
cells must be returned
to the place of purchase or to
a place where they can be
disposed of properly.
Batteries and O
2
cells must
not be disposed of with
ordinary waste.
Important
Pb
Содержание Servo Ventilator 300
Страница 4: ...4 Siemens Elema AB E382 E380E 061 01 03 01 Servo Ventilator 300 300A Notes Notes ...
Страница 6: ...6 Siemens Elema AB E382 E380E 061 01 03 01 Servo Ventilator 300 300A Notes Notes ...
Страница 18: ...20 Siemens Elema AB E382 E380E 061 01 03 01 Introduction Servo Ventilator 300 300A 1 Notes ...
Страница 22: ...20 Siemens Elema AB E382 E380E 061 01 03 01 Introduction Servo Ventilator 300 300A 1 Notes ...
Страница 114: ...5 112 Siemens Elema AB E382 E380E 061 01 03 01 Troubleshooting Servo Ventilator 300 300A Notes ...
Страница 130: ......