68
Siemens-Elema AB
E382 E380E 061 01 03 01
Description of functions
Servo Ventilator 300/300A
2
11 Inspiratory channel
The fresh gas from the
INSPIRATORY
VALVES
(8
and 9) is mixed and lead to the patient
system via this channel. The O
2
cell and the
safety valve are also parts of the inspiratory
channel.
11.1 Inspiratory mixing part
The different fresh gases come together in
this part. It also contains a fitting which
connects to the main block 12
INSPIRATORY
PRESSURE
.
The
INSPIRATORY
MIXING
PART
is designed to
suppress pressure oscillates from going to
the pressure transducer.
11.2 Inspiratory pipe
This plastic pipe leads the mixed gas from
the 11.1
INSPIRATORY
MIXING
PART
to the patient
system. It is held in position by a snap lock.
It contains a holder for the 11.3
O
2
CELL
and it
is integrated with the 11.4
SAFETY
VALVE
seat.
The outlet of this inspiratory pipe is called
INSPIRATORY
OUTLET
.
11.3 O
2
cell
The O
2
cell is mounted in a holder on the
11.2
INSPIRATORY
PIPE
and protected by a
bacteria filter. Service interval 3000 h of
operation for exchange of the O
2
BACTERIA
FILTER
.
The O
2
cell gives an output voltage
proportional to the partial pressure of oxygen
inside the
INSPIRATORY
PIPE
. At constant
pressure this output is proportional to the O
2
concentration in percent. In each O
2
cell the
output signal will stay at a fairly constant
level usually within 1017 mV in normal air
and standard barometric pressure during the
life time of the cell. This signal is used in the
block 15.3
O
2
CONCENTRATION
AMPLIFIER
. The cell
should be replaced when the output level
has started to decrease. In this situation,
repeated calibration of the O
2
concentration is
required.
The life time of the cell is affected by the O
2
concentration. With the O
2
concentration (at
the cell) in % and expected cell life in hours
the following applies at 25°C (77°F):
O
2
concentration x Expected cell life
N
800 000% hours.
A separate output signal (
O
2
CELL
CONNECTED
L
)
indicates if the O
2
cell is connected or not.
This signal is used in the main block 4
MONITORING
. The O
2
cell cable is connected to
the board PC 1607
PNEUMATIC
INTERCONNECTION
(connector P22) inside the
PATIENT
UNIT
.
Содержание Servo Ventilator 300
Страница 4: ...4 Siemens Elema AB E382 E380E 061 01 03 01 Servo Ventilator 300 300A Notes Notes ...
Страница 6: ...6 Siemens Elema AB E382 E380E 061 01 03 01 Servo Ventilator 300 300A Notes Notes ...
Страница 18: ...20 Siemens Elema AB E382 E380E 061 01 03 01 Introduction Servo Ventilator 300 300A 1 Notes ...
Страница 22: ...20 Siemens Elema AB E382 E380E 061 01 03 01 Introduction Servo Ventilator 300 300A 1 Notes ...
Страница 114: ...5 112 Siemens Elema AB E382 E380E 061 01 03 01 Troubleshooting Servo Ventilator 300 300A Notes ...
Страница 130: ......