3
90
Siemens-Elema AB
E382 E380E 061 01 03 01
Disassembling and assembling
Servo Ventilator 300/300A
Disassembling the control
section
PC boards PC 1587, PC 1605,
PC 1608 and PC 1616
• Remove the screws (1) and lift off the
control section lid (2).
• To remove any of the boards PC 1605 (3),
PC 1608 (4) or PC 1616 (5), grasp the
PC board and pull it out.
• To remove PC 1587 (6), first remove
PC 1608 as mentioned above. When
PC 1608 is removed, grasp PC 1587 and
pull it out.
Note
– Before exchanging COM-PROMs
on PC 1587, always disconnect the power
supply from the battery on PC 1587.
See information regarding power supply
disconnection in chapter ”4. Service
procedures”, section ”
data memory and real time clock
6
4
5
3
2
1
300-A33X