MOBILETT Plus / - E/ - M
SPR8-215.898.01
Page 4 of 6
Siemens-Elema AB
Rev. 05
01.02
SPS-UD
Solna, Sweden
7 - 4
Capacitor bank
•
Remove the main bars (Fig. 5 for MOBILETT Plus/ Plus M and Fig. 6 for MOBILETT
Plus E).
•
Remove the capacitor module.
Fig. 5
MOBILETT Plus/ Plus M
Fig. 6
MOBILETT Plus E
Fig. 7
Residual voltage up to 353 VDC may still be present.
Risk of high voltage.
Use insulated tools and do not touch bars that may contain volt-
age (Fig. 7).
MOB00496
+ bar (pos)
- bar (neg)
MOB00495
+ bar (pos)
- bar (neg)
WARNING