MOBILETT Plus / - E/ - M
SPR8-215.898.01
Page 18 of 28
Siemens-Elema AB
Rev. 05
01.02
SPS-UD
Solna, Sweden
10 - 18
Mechanical systems
Replacement of the support arm cable harness
10
•
Position the unit with the single tank resting on a table. Secure the support arm with a
piece of rope (Fig. 21).
•
Remove covers "A" and "B".
•
Remove collimator cover "C".
Fig. 21
Never work with the system open if the capacitor is charging or
has just been charged.
The capacitor bank may still be charged even if the system is
switched OFF and the mains cable is disconnected.
Life-threatening electric shock hazard exists.
See Chap.:
"Protective measures when working on the capacitor bank".
Equipment required:
Rope
MO
C
A
B
DANGER