7 - 1
Siemens-Elema AB
SPR8-215.898.01
Page 1 of 6
MOBILETT Plus / - E/ - M
Solna, Sweden
Rev. 05
01.02
SPS-UD
Capacitor bank
7
Malfunctions detected from the generator:
Leakage current too high
The leakage current increases when the capacitor bank has
been discharged for a long time. The generator displays a
corresponding message:
•
Perform a formation according to section
"P06 - Capacitor formation".
Capacitance too low
If a capacitor malfunctions, the internal series fuse will blow
and the capacitor series will be disconnected.
Depending on the tolerance in the capacitors, the generator
will display a message. If a large positive tolerance is
present in the capacitors, the balance of the capacitor bank
will be able to maintain the total capacitance above the lower
limit. Both the capacitor and the fuse must be replaced
according to the following.
•
Switch off the generator and disconnect the power plug.
Dangerous residual voltage may still be present in the capacitor
bank.
Risk of high voltage.
Consult the safety information in chapter 1.
WARNUNG
WARNING