Display messages
2 - 9
Siemens-Elema AB
SPR8-215.898.01
Page 9 of 22
MOBILETT Plus / - E/ - M
Solna, Sweden
Rev. 05
01.02
SPS-UD
Err 18
Generator ON
The kV value
changes via the
kV+ or kV- but-
ton.
The change in volt-
age level in the
capacitor bank takes
too long
(>0.5 s/V). Refer
also to "Err 17".
•
Check whether S10 on D7 is
switched off.
•
Check if any of the capacitors in the
capacitor bank are defective. The
output energy may be consumed
by a defective capacitor, refer to
section " Capacitor bank".
•
Check the signals
"CHARGE_HIGH", "DISCHARGE"
and "CHARGE_LOW", refer to
page 130**. Check fuse U1 on D7,
U5 and U8 on D8 and resistors
R100 and R101.
CAL 19
Generator goes
ON before the
ready lamp
glows.
The leakage current
in the capacitor bank
is too large.
The measurement is performed dur-
ing charging at 150 V DC. Charging
is interrupted for 5 s.
If the voltage drops more than 1.6 V,
a message will be displayed.
See also "Err 17".
•
In cases where the generator has
not been operated for several
months, the leakage current in the
capacitors has increased. Use the
formation program P06, refer to
section "P06 - Capacitor
formation".
•
If the unit was in regular use before
the message occurred, the error
may result from increased leakage
current or a damaged capacitor,
refer to the section "Capacitor
bank".
•
Check the capacitance in the
capacitor bank, refer to the section
"P03 - Capacitor bank and
charging test". Refer also to page
130**.
Dis-
played
messages
Mode of
operation
Cause
Possible reason
* See section "Monitoring the kV- and mA-controls"
** Wiring diagram X037E
Tab. 1