Dose calculator
MAMMOMAT Balance
SPB7-115.840.01 Page 42 of 62
Siemens AG
Troubleshooting Guide
Rev.02 10.04
Medical Solutions
System configuration for calibration
•
Compression paddle must be in
•
Reference dosimeter must be placed under the compression paddle
•
Dosimeter reference plane must be located at 550mm from focal spot
•
Characteristics, response linearization, and unit of measure of reference Dosimeter must be
configured before to start calibration
•
Reference dose outside kV range of reference Dosimeter must be set at ZERO
Radiation table
To view and edit the radiation table for the selected mammo
press the "Edit"-"Radiation table" menu.