Troubleshooting
MAMMOMAT Balance
SPB7-115.840.01 Page 13 of 62
Siemens AG
Troubleshooting Guide
Rev.02 10.04
Medical Solutions
Troubleshooting
o
If unit doesn’t switch ON refer to dedicated section of this manual.
o
If unit can be switched ON but: no Frame and no Icons appear on the screen and CPU reset
sequence is wrong , check voltages on CPU motherboard 03-188
o
If unit can be switched ON but: Frame and no Icons appear on the screen and CPU reset
sequence is wrong , check CPU 01-170.
o
If unit can be switched ON but: no Frame and no Icons appear on the screen and CPU reset
sequence is correct check keyboard 92-083, KEYBOARD CPU 03-181, cable CV0155 and
supply voltages onboard of KEYBOARD 92-083
If the unit is operative fault finding and troubleshooting is based on the followings:
1) Operating or technical errors on display
2) Technical menu
3) Technical errors in Log file
1) Operating errors on Display
Operating errors on Display are described in the Operator’s manual SPB7-115.620.01….
For the technical aspects related to such alarms:
ALARM
GAS SPRING DEFECTIVE !
Code
Gs
When
Any time when unit is ON
Why
Gas spring has loosed it’s capacity to balance c-arm suspended masses
Action
For Gas Spring replacement refer to “Gas spring replacement” section in this document
For safety reasons unit can’t be operate d if Gas Spring is defective.
Alarm can’t be cleared
WARNING