
Rotating anode starter
MAMMOMAT Balance
SPB7-115.840.01 Page 24 of 62
Siemens AG
Troubleshooting Guide
Rev.02 10.04
Medical Solutions
•
Calibration
During tube running measure Vdc between TP1 and GND
(ground) , then adjust RV1 to have Vdc at
TP3 = 55% of Vdc at TP1.
Repeat same operation with
TP2
,
RV2
and
TP4.
Typical values of
running voltages at
TP1
and
TP2
for IAE tube XM12 are :
IAE
VTP1
1.2 V
VTP2
2.5 V
Factory calibration voltages at TP1, TP2, TP3, TP4 can be found in the tube calibration report.
V(TP3) = 0.55xV(TP1)
V (TP4) = 0.55 x V (TP2)