High Voltage generator
MAMMOMAT Balance
SPB7-115.840.01 Page 34 of 62
Siemens AG
Troubleshooting Guide
Rev.02 10.04
Medical Solutions
kV risetime is typically 1.5 ms in any load condition
Non invasive kV check
kV measurement with non-invasive instruments is affected by beam filtration and MUST be done
exclusively with MOLYBDENUM filter and without compression plate between x-ray source and kVp
meter.
kVp meter calibration factors, non linearity and filter thickness must be taken into consideration to
correct measured values.
kV check at TP1/TP2 PCB 94-132
H.V. generator has voltage divider on which kV check can be done by means of normal DVM with
1000 Vdc full scale or oscilloscope with probe able to withstand a minimum of 500 Vdc.
Connect DVM or oscilloscope to TP1/TP2 (GND) of PCB 94-132.
In the table below kV vs. VTP1 are given.
kV
VTP1 (
±
1%)
20 kV
279 Vdc
21 kV
293 Vdc
22 kV
307 Vdc
23 kV
320 Vdc
24 kV
334 Vdc
25 kV
348 Vdc
26 kV
362 Vdc
27 kV
376 Vdc
28 kV
390 Vdc
29 kV
404 Vdc
30 kV
418 Vdc
31 kV
432 Vdc
32 kV
446 Vdc
33 kV
460 Vdc
34 kV
474 Vdc
35 kV
488 Vdc