Troubleshooting
MAMMOMAT Balance
SPB7-115.840.01 Page 19 of 62
Siemens AG
Troubleshooting Guide
Rev.02 10.04
Medical Solutions
Fault
CPU fuses F2 or F4
Action
Replace fuse
Fault
Filament 92-073 >> FAIL <<
Action
1) check LED status of PCB 92-073
2) if problem persists after having switched the Unit off and on again, verify the integrity of
fuses F1 F2 PCB 92-073 and of filaments by an ohmmeter.
3) if Alarm Code is T2 it was generated during preparation, check for eventual corruption
of filament calibration values
Check also calibration of Over-current and Over-voltage protection circuits.
4) if Alarm Code is R2 it was generated during exposure check calibration of Over-current
and Over-voltage protection circuits.
5) if in doubt that filament board safety circuits are not recognized by CPU, they can be
simulated by means of jumpers on the flat cable CV0173.
Fault
Rotating anode 92-074 >> FAIL <<
Action
1) check LED status of PCB 92-074
2) if problem occurs when the unit is in stand by check that flat cable CV174 connecting
PCB 92-074 and PCB 01-170 has been correctly seated. Also check fuses F1 and F3 of
PCB 92-074
3) if problem occurs in preparation error code is T1 check for line voltage, protection
circuit calibration of PCB 92-074, voltage across shift capacitors C5 and C6 that typically
is about 120 Vac ( 50hz) during anode running. If voltage is lower shift capacitor is
probably defective. For capacitor voltage check, protection circuit has to be disabled.
4) if problem occurs during exposure, error code is R1 check the same as above point 3
5) alarm can also occur due to tube overheating when internal thermo switch opens.
6) if in doubt that starter board safety circuits are not recognized by CPU, they can be
simulated by means of jumpers on CN5 PCB 01-170 or flat cable CV0174.
CN5