72
CLINITEK Analyzer Operator’s Guide
Troubl
eshooting
E59
Inverted strip
positioned on the
test table
Repeat the test with a fresh strip and
ensure that the strip is correctly
positioned on the test table (see
Section 2,
E60
Tilted strip
Repeat the test with a fresh strip and
ensure that the strip is correctly
positioned on the test table (see
E61
Dry strip
Repeat the test with a fresh strip and
ensure that the strip has been in
contact with the sample (see Section
2,
E62
Light Ingress
Too much light is reflecting on the
analyzer. Move the analyzer to a
location with lower lighting.
Contact your local technical support
provider or distributor.
E63
Failure to find end
of strip
Repeat the test with a fresh strip and
ensure that the strip is correctly
positioned on the test table (see
Section 2,
Error
Code
Error Message
Action