1. Select
START
.
The
Prepare Test
screen displays steps on how to perform the test.
A timer displays the amount of time remaining to complete the
task.
2. Draw the urine sample to the line marked on the pipette
(approximately 0.2 mL).
For instructions on the cassette, see the cassette instructions for
use insert.
3. Add the entire contents of the pipette into the sample well of the
test cassette.
After the 8-second countdown ends, the analyzer pulls in the test
table and cassette, and then calibrates.
Note
Each time you run a test, the analyzer calibrates.
After the calibration finishes, the analyzer starts analyzing the
cassette, and the
Analyzing
screen displays.
The analyzer displays either of the following screens:
•
Analyzing
If the cassette is still being analyzed.
•
Results
If analyzing the cassette has been completed.
A timer counts down the time remaining in the cassette analysis
process. After the countdown ends, the analyzer displays the test
results on the
Results
screen.
The results display on the screen for 2 minutes. Then, the display
returns to the
Select Ready
screen.
The test table and cassette move out of the analyzer.
CAUTION
Do not push or pull the test table because the calibration
might fail or the movement might cause table positioning
errors.
Do not move or bump the table while the analyzer calibrates.
The calibration might fail.
Содержание CLINITEK Status+
Страница 28: ...28 CLINITEK Status Analyzer Operator s Guide ...
Страница 50: ...50 CLINITEK Status Analyzer Operator s Guide Operations ...
Страница 56: ...56 CLINITEK Status Analyzer Operator s Guide Calibration QC ...
Страница 84: ...84 CLINITEK Status Analyzer Operator s Guide File Management ...
Страница 128: ...128 CLINITEK Status Analyzer Operator s Guide System Configuration ...