
CLINITEK Analyzer Operator’s Guide
123
System C
onf
iguration
To upgrade the analyzer software, perform the following steps:
1. If the CLINITEK analyzer is on, press the on/off power
button until the analyzer powers off.
The test table retracts.
2. Prepare the analyzer:
a. Ensure the CLINITEK analyzer connects to external
power and not battery power.
b. Do not power on the analyzer.
c. Turn the analyzer so that the back of the analyzer faces you.
d. Lift the printer cover.
CAUTION
Ensure you have printed or recorded the most recent patient
results before you perform the software upgrade because the
upgrade process deletes all patient records and all patient test
results in the system. For more information about recall
results, see Section 6,
CAUTION
Do not use battery power when you upgrade the software. If
you do, the software installation might fail.
CAUTION
Do not use gloves when you insert or remove the memory
card. Using gloves can result in electrostatic damage to the
card.
Ensure you wear gloves as required by your facility to
perform other tasks.
Содержание CLINITEK Status+
Страница 28: ...28 CLINITEK Status Analyzer Operator s Guide ...
Страница 50: ...50 CLINITEK Status Analyzer Operator s Guide Operations ...
Страница 56: ...56 CLINITEK Status Analyzer Operator s Guide Calibration QC ...
Страница 84: ...84 CLINITEK Status Analyzer Operator s Guide File Management ...
Страница 128: ...128 CLINITEK Status Analyzer Operator s Guide System Configuration ...