70
CLINITEK Analyzer Operator’s Guide
Troubl
eshooting
E27
Setup failure
1. Power off the analyzer by
pressing the on/off button for 2
seconds.
2. Power on the analyzer by
pressing the on/off button.
E28
Printer error
1. Lift the printer cover.
2. Push the paper holding arm back
into position.
For the location of the paper holding
arm, see Section 1,
E50
Incorrect strip type
or tilted strip
Note
For ID band urinalysis strips,
skip step 1.
1. Ensure that the strip type you
selected in Instrument Set Up is
the type you use (see Section 7,
2. Verify that you correctly placed
the strip on the test table insert.
If you used the correct type of
strip and you correctly placed
the strip on the test table
insert, check the analyzer
operation by running either of
the following tests:
○
Test a yellow and clear
sample.
○
Run a Chek-Stix QC test (see
Section 3,
E52
Invalid barcode
Repeat the test using the correct
Siemens cassette.
Error
Code
Error Message
Action