114
CLINITEK Analyzer Operator’s Guide
System C
onf
iguration
•
Select a selected chemistry to exclude it.
6. Select
Next
to display the
Reported Chemistries 2 of 2
screen,
which contains the rest of the chemistries.
•
Select a chemistry to include it.
•
Select a selected chemistry to exclude it.
7. To return to the
Select Ready
screen, perform the following steps:
a. Select
Done
.
b. Select
Next
.
c. Select
Done
twice.
Setting up the Authorized Operators
The CLINITEK analyzer stores 700 operators. You can authorize
operators to perform certain tasks, where they gain access to the
system by entering their name to perform those tasks. You can add,
edit, and delete authorized operators.
Note
When you enable the Instrument Set Up password, you restrict
access to Instrument Set Up for only those operators who know the
password. If you enable the Authorized Operator setting and the
Instrument Set Up password, the operator name has priority over the
Instrument Set Up password.
Enable or Disable Authorized Operators
To enable or disable authorized operators, perform the following steps:
1. On the
Select Ready
screen, select
Instrument Set Up
.
2. On the
Choose Settings
screen, select
Instrument Settings
.
3. On the
Instrument Settings
screen, select
Authorized Operator
.
4. On the
Authorized Operator
screen, for
Authorized operators
only
, select an option:
•
To permit access only by authorized operators, select
Enabled
.
Содержание CLINITEK Status+
Страница 28: ...28 CLINITEK Status Analyzer Operator s Guide ...
Страница 50: ...50 CLINITEK Status Analyzer Operator s Guide Operations ...
Страница 56: ...56 CLINITEK Status Analyzer Operator s Guide Calibration QC ...
Страница 84: ...84 CLINITEK Status Analyzer Operator s Guide File Management ...
Страница 128: ...128 CLINITEK Status Analyzer Operator s Guide System Configuration ...