42
8024638/AE00/V1-0/2019-09| SICK
S E R V I C E M A N U A L | MCS200HW
Subject to change without notice
3
MAINTENANCE
3
Place the crystal glass pane on the tube.
4
Place O-ring 3 on the crystal glass pane.
Fig. 52: Positioning tube O-rings and crystal glass.
6
Place the lower mirror plate over the tube from above.
Fig. 53: Lower mirror plate with tube
7
Turn the mirror plate with the tube so that the tube points upwards.
Fig. 54: Aligning the tube on the mirror plate
8
Turn the tube to align it so that the holes for the screws are over each other. The tube
must end flush with the mirror plate.
9
Screw the tube tight onto the lower mirror plate.
10
Repeat procedure for second tube.
11
Fasten the lower mirror plate to the cell with a new O-ring (
rings of the lower mirror plate”, page 37
12
Refit the cell (
see Chapter “Fitting the complete cell”, page 18
Check new crystal glass for damage before positioning.
SIC_18_MCS200-033.tif
Prepare screws with washers for the tubes.