70
8024638/AE00/V1-0/2019-09| SICK
S E R V I C E M A N U A L | MCS200HW
Subject to change without notice
5
REPAIRS
8
Screw the lower cover plate tight.
Fig. 94: Cell, lower cover plate
9
Position the tubes O-rings.
10
Fasten the hose lines with fastening clips.
11
Insulate the non-return valve with Teflon tape and screw tight (
the non-return valve on the cell inlet filter”, page 25
Fig. 95: Insulating the non-return valve
5.2.3
Replacing the oxygen sensor
The cell inlet filter can be replaced with the cell fitted.
1
Fastening screw
2
Sealing ring
Tools required
Phillips screwdriver
Spare parts set No.
Contents
2081837
Oxygen sensor with double-end ring spanner
DANGER: Risk of burns due to hot surface.
Allow the MCS200HW to cool down before carrying out any work.