REPAIRS
5
Fig. 117: PT100
1
Remove the complete cell (
see Chapter “Removing the complete cell”, page 65
2
Place the cell down with the rear side upwards.
>>> The side with the cell inlet filter and ejector is at the bottom.
Fig. 118: Cell, overview, left
3
Loosen the fastening clips with a Phillips screwdriver (2 pieces).
4
Remove the PT100.
5
Remove the PT100 cable from the cable conduit.
1
Open the contact plug.
▸
Carefully lever open the clamps of the contact plug with a slotted screwdriver
(3 pieces).
Fig. 119: Opening the contact plug
1
Fastening clips
2
PT100