63
8024638/AE00/V1-0/2019-09| SICK
S E R V I C E M A N U A L | MCS200HW
Subject to change without notice
REPAIRS
5
Fig. 78: Display, fastening clips
4
Push the fastening clips to the side.
5
Replace the display.
6
Tighten the fastening clips again.
7
Reconnect the cables.
8
Connect the MCS200HW power supply.
9
Parameterize the IP addresses of the display.
5.1.8
Replacing the 24V power supply unit
1
Disconnect the MCS200HW power supply.
Fig. 79: Device cabinet, 24V power supply unit
1
Fastening clip
Tools required
Phillips screwdriver
Spare parts set No.
Contents
2099745
Power supply unit 85-264 V, 24 V, 2.5 A
Observe the safety rules for work on electrical systems (
).
1
24V power supply unit