65
8024638/AE00/V1-0/2019-09| SICK
S E R V I C E M A N U A L | MCS200HW
Subject to change without notice
REPAIRS
5
5.2
Working on the cell
5.2.1
Removing the complete cell
1
Remove the cell (
see Chapter “Removing the complete cell”, page 15
).
2
Unscrew the 90-degree bracket of the test gas line on the cell inlet filter using a jaw
wrench (
see Fig. 6 “Cell open”, page 16
3
Place the cell down with the upper side facing downwards.
4
Remove the A-holder on both sides.
▸
Loosen the fastening screws with a 5 mm Allen key (3 pieces).
Fig. 82: Cell, A-holder
DANGER: Risk of burns due to hot surface.
Allow the MCS200HW to cool down before carrying out any work.
Tools required
Phillips screwdriver
5 mm Allen key
Jaw wrench
The non-return valve must then be insulated with Teflon tape.
1
Fastening screw