40
8024638/AE00/V1-0/2019-09| SICK
S E R V I C E M A N U A L | MCS200HW
Subject to change without notice
3
MAINTENANCE
3.4.6.3
Tubes maintenance
The complete cell must be removed for tube maintenance (
).
1
Loosen the fastening screws of the tubes with a 3 mm Allen key (3 pieces each).
Fig. 48: Tubes
2
Lever out the tubes:
1
Screw the screws into the thread openings provided (2 pieces).
Fig. 49: Thread openings, tubes
2
Tighten the fastening screws with a 3 mm Allen key:
▸
Tighten fastening screws parallel.
▸
Ensure the screws grip properly and avoid thread seizure
3
Take off the tubes.
Carry out tube maintenance at the same time as maintenance of the lower mirror plate.
Tools required
3 mm Allen key
5 mm Allen key
1 Fastening screws, tube
Keep the screws for reassembly.
1 Thread opening with screw
2 Thread openings, tubes