47
8024638/AE00/V1-0/2019-09| SICK
S E R V I C E M A N U A L | MCS200HW
Subject to change without notice
FUNCTIONAL TEST
4
4.3
Beamer current check
1
Select
Diagnosis
->
State control
->
Beamer
.
2
Check beamer current. The beamer current must be within the specified warning
thresholds.
3
Replace the beamer when the radiator current is less than 0.05 A above the
Current
Min. Warning
or a maintenance request is pending.
4.4
Reference energy check
1
Select
Diagnosis
->
Control values
->
Reference energy
.
2
Check the reference energy.