18
8024638/AE00/V1-0/2019-09| SICK
S E R V I C E M A N U A L | MCS200HW
Subject to change without notice
3
MAINTENANCE
3.4.2
Fitting the complete cell
1
Guide the power supply for the cell heater through the recess in the device cabinet.
Fig. 10: Device cabinet with retaining plate
2
Position the cell in the device cabinet:
▸
Hook the bolts of the bracket plate of the cell into the retaining plate in the device
cabinet.
Fig. 11: Cell with bracket plate
3
Screw the cell A-holder tight:
▸
Fasten screws left and right with a 5 mm Allen key (3 pieces each).
Tools required
Jaw wrench, size 14/16/18
Phillips screwdriver
5 mm Allen key
1 Retaining plate
2 Recess
3 Power supply, cell heater
1 Bracket plate with bolts