5.2.5.2
Checking the data archives (data logs)
FLOWSIC600-XT has a diagnostics archive and two data archives:
● Data archive 1 (Hourly Log) on an hourly basis
● Data archive 2 (Daily Log) on a daily basis
The measured data archives are saved accordingly in the non-volatile memory of the
device.
All data archives can be loaded and read in the “Archive Management” overview. In the
individual archives, each archive can be loaded individually from the device to the PC.
The archive data can be exported in CFX or CSV Format and then stored or sent as e-mail.
Fig. 53
Example: Data archive 1
Deleting the data archives
The data archives can be deleted via the FLOWgate™ operating software. In the archive
overview of the “Archive Management”, all data archives can be deleted at once or each
individually directly in the respective archive.
5.2.6
Creating and evaluating the Diagnostics Comparison report
FLOWgate supports the creation of a Diagnostics Comparison report. This visualizes a
comparison of the most important current process and diagnostic data with the fingerprint
data stored in the device during commissioning. The data is displayed for the individual
speed classes.
The report can be created at any time and stored as a PDF document. This enables, for
example, trend reporting over the lifetime of the meter with which process or meter
changes can be detected.