124
FLOWSIC600-XT · Operating Instructions · 8018846/1BHP/V2-1/2021-05 · © SICK Engineering GmbH
Specifications
Su
bje
ct to
ch
an
ge
w
ith
ou
t n
otic
e
8.3
Measuring ranges
Table 19
Measuring ranges (metric)
Only the “Standard flow range according to MID" is applicable for device version FLOWSIC600-XT C.
The maximum gas velocity in the pipe is limited to 40 m/s in configurations with flow conditioner.
H[WHQGHG
MID
PLQLPXP
IORZUDWH
[m³/h]
Standard
MID
PLQLPXP
IORZUDWH
[m³/h]
MID
WUDQVLWLRQ
IORZUDWH
[m³/h]
MID
PD[LPXP
IORZUDWH
[m³/h]
N
RQ
-MID
PD[LPXP
IORZUDWH
[m³/h]
0HWHUVL]H
H[WHQGHG
Q
min
Standard Q
min
DFF
. ISO17089
Q
t
Standard
Q
max
e
[WHQGHG
Maximal Q
max
V
max
[3]
V
t
DN80 (3")
5
8
40
650
1.000
61
2,5
DN100 (4")
8
13
65
1.000
1.600
63
2,5
DN150 (6")
16
20
100
2.
00
3.000
52
1,7
DN200 (8")
20
32
160
4.000
4.500
44
1,6
DN250 (10“)
25
50
240
6.
00
7.000
44
1,5
DN300 (12")
35
65
310
7.800
8.000
39
1,5
DN350 (14")
45
80
420
10.000
10.000
36
1,5
DN400 (16")
60
120
550
13.000
14.000
38
1,5
DN450 (18")
100
130
700
16.000
17.000
37
1,5
DN500 (20")
130
200
850
20.000
20.000
35
1,5
DN550 (22“)
150
260
1.000
24.000
24.000
35
1,5
DN600 (24")
180
320
1.200
28.000
32.000
39
1,5
DN650 (26“)
240
450
1.400
32.000
35.000
36
1,5
DN700 (28")
280
650
1.700
36.000
40.000
36
1,5
DN750 (30")
320
650
1.900
40.000
45.000
35
1,5
DN800 (32")
360
800
2.200
43.000
50.000
34
1,5
DN850 (34“)
400
900
2.500
47.000
55.000
33
1,5
DN900 (36")
450
1.000
2.800
51.000
66.000
36
1,5
DN950 (38")
500
1.100
3.100
56.000
70.000
34
1,5
DN1000 (40")
550
1.200
3.400
60.000
80.000
35
1,5
DN1050 (42")
600
1.300
3.800
65.000
85.000
34
1,5
DN1100 (44")
650
1.400
4.100
70.000
90.000
32
1,5
DN1150 (46")
700
1.500
4.500
72.000
95.000
34
1,5
DN1200 (48")
750
1.600
4.800
80.000
100.000
30
1,5
DN1300 (52")
900
1.700
5.600
90.000
110.000
28
1,5
DN1400 (56")
1.000
1.800
6.500
100.000
120.000
27
1,5
6WDQGDUGIORZUDWHUDQJHDFF0,'
([WHQGHGIORZUDWHUDQJHDFF0,'
m/s
9HORFLW\