2.9.2
Archives
● One configurable diagnostics archive (6,000 entries)
Diagnostics data are saved in cyclic intervals in the diagnostics archive. The storage
period can be set in the range from 15 min to 6 h. Dataset number, timestamp, various
global measured values, status information and path information are saved. The archive
has a capacity of 6000 entries and is circulating by default. The archive serves primarily
for analysis of historical measured data.
● Two configurable archives (6,000 entries each)
Accounting data are saved in cyclic intervals in data archives 1 and 2. The storage
period can be set in the range from 15 min to 24 h. Dataset number, timestamp, status
information, various meter levels as well as various operating variables and
standardized variables are saved. The archives have a memory of 6000 entries and are
circulating by default.
Table 2
Data archive contents and structure
Element
Significance
Date record number
Consecutive number of the data record, is not reset when the logbook is cleared.
Timestamp
Time of the entry as Unix timestamp (UTC)
Unit ID
Bits
0 : Pressure type (0=absolute, 1=relative)
1 : Unit system (0=metric, 1=imperial)
2..4 : Pressure unit
5..7 : Temperature unit
Flowtime
Percentage of periods in which the throughflow was in the recording direction [%]
Detail status
Detailed status information (ActualStatus)
Meter 1 : V
Totalizer 1 : Volume uninterrupted/total
Meter 1 : Verr
Totalizer 1 : Volume when meter in malfunction status
Meter 1 : ID
Totalizer 1 : Bits
0 : Totalizer status (0=uninterrupted, 1=total)
1..2 : Totalizer type (0=operation, 1=standard, 2=mass, 3=reserved)
3..6 : Power of ten totalizer resolution plus 8
7 : Unit system (0=metric, 1=imperial)
Meter 2 : V
Totalizer 2 : Volume uninterrupted/total
Meter 2 : Verr
Totalizer 2 : Volume when meter in malfunction status
Meter 2 : ID
Totalizer 2 : Bits
0 : Totalizer status (0=uninterrupted, 1=total)
1..2 : Totalizer type (0=operation, 1=standard, 2=mass, 3=reserved)
3..6 : Power of ten totalizer resolution plus 8
7 : Unit system (0=metric, 1=imperial)
Pressure
Pressure (average
[1]
of the measuring period)
[1]
) The values are throughflow-weighted in the recording direction when the throughflow was within the period in
recording direction. The values are averaged when no throughflow existed within the period.
Temperature
Temperature (average
(1)
of the measuring period)
Compressibility
Compressibility (average
(1)
of the measuring period)
Conversion factor
Conversion factor (average
(1)
of the measuring period)
SOS
Sound velocity (average
(1)
of the measuring period)
Molar mass
Molar mass (average
(1)
of the measuring period)
Density
Density (average
(1)
of the measuring period)
Reserved
Field reserved (for future extensions (must be zero!)
Check sum
CRC-16 check sum for the data record