Operating Instructions
CLV 480 Bar Code Scanner
I-4
©
SICK AG · Division Auto Ident · Germany · All rights reserved
8 010 080/O824/10-02-2005
Quick Finder
Installation procedure (overview)
CLV in stand-alone configuration (without heater)
Start/stop mode: Reading trigger via “Sensor“ switching input (default setting)
1.
Check the delivery to make sure that none of the components is missing.
2.
Mount the CLV at the reading station and align it with the object carrying the bar code.
3.
Mount the AMV/S 60 Connection Module.
4.
Connect the CLV to the AMV/S 60 Connection Module using two cables no. 2 020 302.
Alternatively, connect the device to the AMV/S 60 via the external parameter memory
no. 2 020 307.
5.
Mount the sensor for starting/stopping the reading pulse.
6.
Connect the reading pulse sensor to the "Sensor" switching input in the AMV/S 60.
7.
Connect the host to the host interface in the AMV/S 60.
Adapt the AMV/S 60 to the host interface type of the CLV.
8.
Switch on the power supply to the AMV/S 60.
The "Device Ready" LED lights up after the CLV has started.
CLV with external empty parameter memory connected:
After the CLV has started, it immediately copies the internal parameter set to the exter-
nal parameter memory if the memory is empty. Then the "Device Ready“ LED lights up.
Line scanner with oscillating mirror:
In the default setting, the CLV deflects the scan line around the position CW = 50 with
a frequency of 1 Hz and an oscillating amplitude of max.
±
20
°
(
±
40 CW).
9.
Switch on your PC and start Windows
TM
(minimum requirement: Windows 95
TM
).
10. Install the "CLV-Setup" software and the online CLV-Setup Help from the CD-ROM
("Manuals & Software") on your PC.
11. Connect the PC to the terminal interface of the CLV.
To do so, connect a 3-core RS 232 data cable (null modem cable), e. g. no. 2 014 054
to the "Service plug" in the AMV/S 60.
12. Start the "CLV-Setup" program.
CLV-Setup establishes communication with the CLV and uploads the parameter set.
The parameters are then displayed on the tabs.
13. Carry out a test read using test bar codes (clock the CLV accordingly).
Display the reading result in the Terminal Emulator window of the "CLV-Setup" program.
14. Configure the CLV for the application using the settings on the tabs in CLV-Setup.
Copy (download) the modified parameter set to the CLV temporarily.
Do
not
switch off the power to the AMV/S 60 (CLV)!
15. Test the application under realistic conditions.
16. Check whether the data is transmitted correctly between the CLV and host.
17. If necessary, correct and optimize the parameter values.
Copy (download) the parameter set
permanently
to the CLV.
CLV with external parameter memory connected:
Copy the modified parameter set to the external parameter memory when CLV-Setup
asks you for confirmation.
18. Save the parameter set as a configuration file "*.scl" in the "CLV-Setup" program.
The CLV can then be operated with the
application-specific
settings.