Operating Instructions
Chapter
8
CLV 480 Bar Code Scanner
8 010 080/O824/10-02-2005
©
SICK AG · Division Auto Ident · Germany · All rights reserved
8-17
Troubleshooting
8.5.4
Malfunctions in Reading mode: errors in the result status output
¾
Check whether the "Device Ready" LED lights up.
If the LED does not light, see
.
CLV in stand-alone configuration
8.5.5
Malfunctions in Reading mode: oscillating mirror errors
¾
Check whether the "Device Ready" LED lights up.
If the LED does not light, see
.
Malfunction
Possible cause
Remedy
The
– "Result 1"
(default setting: Device Ready)
– "Result 2"
(default setting: Good Read)
– "Result 3"
(default setting: No Read)
– "Result 4"
(default setting: Match 1)
switching outputs are not outputting any
pulses
•
The event associated with the assigned
function for outputting the result status of
the read does not occur in the reading
operation
•
Incorrect switching mode parameterized
in the CLV
•
Not applicable.
•
In the "CLV-Setup" program: choose the
D
EVICE
C
ONFIGURATION
tab. Click the
R
ESULT
O
UTPUT
P
ARAMETERS
button. Are
the outputs set to N
O
I
NVERT
O
UTPUT
?
Change if necessary. Download to CLV!
Table 8-10: Troubleshooting: errors in the result status output in Reading mode
Malfunction
Possible cause
Remedy
1. The oscillating mirror does not move.
Incorrect operating mode set on the CLV
In the "CLV-Setup" program choose the
O
SCILLATING
M
IRROR
tab, O
SCILLATING
M
IRROR
section.
Is O
SCILLATING
WITH
M
AXIMUM
A
MPLITUDE
or
O
SCILLATING
WITH
V
ARIABLE
A
MPLITUDE
selected?
2. The oscillating mirror oscillates with
maximum deflection (40 CW), even
though a variable amplitude, e. g. 10 CW,
is set in the distance configuration 1
•
O
SCILLATING
WITH
M
AXIMUM
A
MPLITUDE
set
on the CLV by mistake
•
The oscillating amplitude is paramterized
in the wrong distance configuration
(here: no. 2)
•
The distance configuration (focus
position) was not changed
•
From the O
SCILLATING
M
IRROR
tab, choose
O
SCILLATING
WITH
V
ARIABLE
A
MPLITUDE
.
Download to the CLV!
•
In the "CLV-Setup" program: choose the
R
EADING
C
ONFIGURATION
tab, click the
D
ISTANCE
C
ONFIGURATION
button.
Is the amplitude value set in the correct
distance configuration (here: no. 1)?
•
In the "CLV-Setup" program: choose the
D
EVICE
C
ONFIGURATION
tab. A
SSIGNMENT
OF
I
NPUTS
section: for IN 3 and IN 4: is dyna-
mic focus control selected? Connect
sensors to the switching inputs. Check
sensor wiring (see
).
Measure output signal at sensor
3. One-Shot: The One-Shot function cannot
be triggered (sensor correctly connected
to the "IN 3" or "IN 4" switching input)
•
Trigger source incorrectly parameterized
•
"IN 3" or "IN 4" switching input has not
been assigned this function
•
In the "CLV-Setup" program: choose the
O
SCILLATING
M
IRROR
tab. Is one-shot sel-
ected?
•
Choose D
EVICE
C
ONFIGURATION
tab. A
SSI
-
GNMENT
OF
I
NPUTS
section: for IN 3 or
IN 4: is trigger One-Shot selected?
Table 8-11: Troubleshooting: oscillating mirror errors in Reading mode