© SICK AG • Auto Ident • Germany • All rights reserved
30
Operating Instructions
8 008 224/I982/08-00
CLV 41x Bar Code Scanner
11. 1 Transferring the application-specific parameter set
If you have to replace a CLV on site and therefore transfer the active parameter set
to the new device, there are two methods of doing so. The parameter set can only
be transferred to a device of the same type.
Transferring the parameter set locally using profile bar codes
Prerequisite:
the active parameter set is available in archived form as printed
profile bar codes.
It is not necessary to connect the PC to the CLV Setup program.
11. Replacing a CLV 41x
Step
Action
1
Disconnect the power supply to the AMV/AMS Connection Module
2
Disconnect the CLV connector from the AMV/AMS Connection Module.
Remove the CLV from the bracket on the reading station.
Mark the position and alignment on the bracket/station.
3
Mount the new CLV and connect it to the AMV/AMS
4Reconnect the power supply to the AMV/AMS Connection Module
5
Present the first archived profile bar code to the CLV within
5 seconds
5 seconds
5 seconds
5 seconds
5 seconds
after the first
acoustic
signal.
The CLV confirms the good read with an acoustic signal and waits 10 s for the
next profile bar code.
6
Present all of the other profile bar codes consecutively
in the order in which they
were printed.
7
Finally, allow the last 10 s wait time to expire.
The CLV then outputs two consecutive acoustic signals and switches to Reading
mode. The Device Ready LED lights up.
The new CLV can now be used with the transferred parameter set.
Table 18
Notes:
The procedure for printing out the profile bar codes is described in
Table 14
in
Section
10.3 Profile programming
.
If none of the profile bar codes of the parameter set for the device to be replaced
has been printed out and if the configuration file (*.scl) has not been saved on the
PC, you can try, depending on the reason why the device is to be replaced, to
upload the parameter set from the CLV to be replaced to the user interface and then
download it to the new CLV.
Ø
Transferring the parameter set by means of a download
If the profile bar codes of the parameter set have not been printed out but a
configuration file (*.scl) does exist, you must connect the PC with the CLV Setup
software to the new CLV (see Section
8.4 Connecting the PC to the CLV
).
In the following section, it is assumed that the new CLV is using the default settings.
10. 6. 5 Operating data
This function allows you to interrogate and reset important counters in the user
interface that are updated continuously by the CLV during the reading procedure.
To open the dialog window, click the icon with the CLV connected. The CLV data
is then loaded to the window directly and displayed.