DATOSPIR
· User’s Manual
EN 26
511-D00-MUM · Rev 1.01
Annex 1. ELECTROMAGNETIC COMPATIBILITY
Guidance and manufacturer’s declaration – electromagnetic emissions
DATOSPIR
is intended for use in the electromagnetic environment specified below. The customer or the user should assure that it is used in such an environment..
Emissions test
Compliance
Electromagnetic environment - Guidance
RF (Radiated) emissions
CISPR 11 (EN 55011)
Grupo 1
Clase B
RF (Conducted) emissions
CISPR 11 (EN 55011)
Not applicable
DATOSPIR
works with batteries.
Harmonic emissions
EN-IEC 61000-3-2
Not applicable
DATOSPIR
works with batteries.
Voltage fluctuations / Flicker emissions
EN-IEC 61000-3-3
Not applicable
DATOSPIR
works with batteries.
Guidance and manufacturer’s declaration – electromagnetic emissions
DATOSPIR
is intended for use in the electromagnetic environment specified below. The costumer or the user of
DATOSPIR
should assure that it is used in such an environment.
Tests EN-IEC 61000-4-4 y -4-5 are applicable to AC/DC power inputs, and input/output signal. Test EN-IEC 61000-4-6 additionally applies to patient connections. Test EN-IEC 61000-4-11
only applies to AC power inputs.
Immunity test
EN-IEC 60601 test level
Compliance level
Electromagnetic environment –Guidance
Electrical fast transient/burst
EN-IEC 61000-4-4
±2 kV for power supply lines and earth
±1 kV for input/output lines
Frecuency 100 kHz
Not applicable
Not applicable
DATOSPIR
works with batteries.
The input/output line cables are shorter than
3 meters long.
Surge
EN-IEC 61000-4-5
±0.5, ±1 kV differential mode (Line to line).
±0.5, ±1, ±2 kV common mode (Line to
ground).
Not applicable
Not applicable
DATOSPIR
works with batteries.
Conducted RF
EN-IEC 61000-4-6
3 Vrms
de 150KHz a 80 MHz
6 Vrms in ISM bands
80% AM a 1kHz
3 Vrms
Voltage dips, short interruptions
and voltage variations on power
supply input lines
EN-IEC 61000-4-11
0% Ut ; 0.5 cycles: 0º, 45º, 90º, 135º, 180º,
225º, 270º y 315º.
0% Ut; 1 cycle
70% Ut; 25/30 cycles: 0º
0% Ut ; 250/300 cycles
(5 segundos)
Not applicable
DATOSPIR
works with batteries.
NOTE Ut is the a.c. mains voltage prior to application of the test level.