DATOSPIR
· User’s Manual
EN 23
511-D00-MUM · Rev 1.01
RANGES AND MEASURES (According to ATS/ERS 2005)
Fleisch
Turbine
Lilly Disposable
Measurement range (BTPS)
Flow
Volume
0 to ±16 l/s
0 to 10 l
Flow resistance 14 l/s (cmH
2
0 / l/s)
<1,4
<0,6
<1,2
Measurement accuracy (BPTS)
Volume
Flow
PEF
(whichever is greater)
3% or 50 ml
5% or 200 ml/s
10% or 300 ml/s (ATS/ERS 2005)
10% or 170 ml/s (EN ISO 23747:2015)
Temporal accuracy
0,50%
Volume resolution
<10ml
<10ml
<10ml
Frequency of sampling
100Hz
100Hz
100Hz
Transducer life
1400 disinfect.
or 3 years
1400 disinfect.
or 3 years
Only one use
(Expir. 3 years)