LABOKLAV Eco 80 / 135
User Manual
19
Sterilization phase has finished, now the chamber
pressure will be reduced until reaching 10 kPa in
chamber.
Pressure reduce is over, now the chamber is venting
until reaching set point.
Venting the chamber finished. An additional time is
running for safety.
The program cycle has finished successfully. The
sterilizer gives the result as correct cycle. Now the
program needs to be quit by pressing the Stop button.
Then the lid can be opened and the sterilization good
can be removed from chamber.
In case of a not successful or broken program the sterilizer shows the following message:
Before regular end of cycle a break was initialized maybe
by hand or automatically. In case of automatic break an
error message is additionally shown.
Cancellation of program cycle
Each program is able to be broken manually by operator by pressing the STOP button. The
program asks if cancellation is really necessary before breaking. After breaking the program has
different possibilities to react. The reaction depends on the situation and the program in that the