LABOKLAV Eco 80 / 135
User Manual
12
6. Use of the autoclave
The steam sterilizer line Laboklav ECO 80 / 135 allows a fully automatic process cycle. That
includes deaeration, heating, sterilization, pressure remove and cooling. All steps for an automatic
process are controlled by a microprocessor control board. The actual status is shown on a graphic
display and includes all important information for the user to operate the device. The supported
temperature range of the sterilization process is 103°C to 136°C. The typical temperatures of
121°C and 134°C are programmed for different sterilizing situations and different materials. All
program positions can be reprogrammed by special trained engineers / service stuff.
The steam sterilizer line Laboklav ECO 80 / 135 includes the following additional advantages:
-
Fully automatic deaeration of the sterilization goods by gravitation process.
-
Simple and easy to use construction of the complete unit
-
No contact of the heaters with feed water safes live cycle time
-
Protection of the drain on house installation side by using normal tape water mixing to the
steam outlet. That process works temperature controlled by PT 100 in the outlet installation.
The Temperature is programmable for opening and closing the mixing valve.
-
Enforced cooling option available (option M): cooling with air
-
Timer controlled program start possible
-
Microprocessor controlled process for fully automatic use
-
Documentation by data readout or printer readout possible
Total view of the steam sterilizer Laboklav ECO xxx
4
5
The Laboklav ECO should be operated while the normal laboratory working hours. If the
device is not in use it should be switched off by the main power plug (5). Over night and on
the weekends the central power plug (1) should be switched off. If that central power plug is
missing please use the main breaker below the housing on the right side of the unit (4).
Emergency switch off
In case of a fatal error switch off the device by turning the central main power switch
(1) or disconnect the sterilizer from power supply by using the power plug (2).
1
2