LABOKLAV Eco 80 / 135     

 

    User Manual 

 

26 

 

        -  Cleaning  the  chamber  inside  –  For  cleaning  the  chamber  use  a  wet  and  soft  towel  from 

cotton material. Special cleaning material or chemicals are not necessary. Do not use aggressive or 

organic chemical for cleaning! Chemicals can damage the sealing or sensors!   

 

 

The  rest  of  chemicals  or  cleaning  materials  will  be  brought 

forward  to  the  sterilization  goods  of  next  sterilization  cycle! 

Do not use aggressive or organic chemicals for cleaning!   

 

 

        - Cleaning the accessories – Clean the baskets etc. with wet towel or under running water.   

 

        - Cleaning the housing – The housing needs to be cleaned by wet towel or light oil. Special 

cleaning chemicals like used in house hold can be used. 

 

Check of the safety valve 

The safety valve(s) needs to be checked once per year. This should be done by specially trained 

service stuff. Other safety checks are necessary so we recommend making one safety check 

together with the yearly maintenance check by a trained service engineer. While testing the function 

of the safety valve steam is leaving the safety valve. 

 

 

Attention please! While steam is leaving the safety  valve the 

valve is getting very hot in a very short time! Make sure that 

your  human  skin  is  not  coming  in  contact  with  the  steam! 

Contact with the steam can burn your skin! 

 

 

 

If the valve is not closing completely after testing it needs to 

be changed! If there are doubts about the regular functionality 

of the tested valve it needs to be changed! 

 

Changing the venting air filter   

Wear  and  tear  of  the  venting  air  filter  depends  on  the  number  of  cycles  and  the  quality  of  the 

environment. We recommend changing the filter after 100 cycles or once per month.   

 

 

 

 

 

 

 

Содержание Laboklav ECO 135

Страница 1: ...Laboklav ECO Steam Sterilizer Basic with Forced Cooling 80 Litres 135 Litres User Manual Laboklav Laboklav Laboklav Laboklav Steam sterilizers ...

Страница 2: ...OKLAV Eco 80 135 User Manual 2 SHP Steriltechnik AG Schloss Detzel 1 D 39345 Detzel Schloss Tel 49 0 39058 97 62 0 Fax 49 0 39058 97 62 22 E Mail info shp steriltechnik de Internet www shp steriltechnik de ...

Страница 3: ...f delivered parts 5 2 Unpacking 5 3 Installation 5 4 Intended Use 6 5 Product specifications 8 6 Use of the autoclave 12 7 Troubleshooting 22 8 Maintenance 24 9 User replaceable accessories and spare parts 27 10 Description of safety devices 27 11 Definition of feed water quality 28 12 Service and maintenance 29 ...

Страница 4: ...an cause injury and even death This symbol also means that an operator must acquaint with a suitable passage in the manual Important This symbol denotes important indications for example to prevent sterilizer or load damage Attention Observing the texts marked with this symbol facilitates operation of the sterilizer General Warnings Access to sterilizer operations manual should be restricted only ...

Страница 5: ...ure comfort in operation while ensuring the minimal distances min 10 cm to the wall and next device in room Due to large amount of heat generated during the sterilizer operation the room in which it is seated should be fitted with a mechanical intake and exhaust ventilation system providing 6 10 exchanges of air per hour The ventilating hood should be installed above the sterilizer By using the tu...

Страница 6: ...ber 4 Intended use The steam sterilizer Laboklav ECO 80 135 is equipped with a chamber by 80 Liter or 135 Liter chamber volume It is constructed for the steam sterilization of instruments materials for non medical use and liquids Liquids should be water based solutions only All versions contain a thermo lock acc to IEC 61010 2 42 Different versions depending on the use of the sterilizers are avail...

Страница 7: ... of acid base or organic solutions It is forbidden to use the autoclave for handling of explosive materials Attention All materials you bring inside the chamber come in contact with water steam It could happen that a chemical reaction will be initiated Sterilizing organic solution may damage the autoclave or reduce the life cycle or maintenance cycle It may be dangerous for the user of the autocla...

Страница 8: ...7 23 EG CE 0036 Cat III Module B C1 Power supply Voltage 1N 230V 5 50 Hz 16A Working power 3 kW Averaged power consumption per cycle 5 kWh Protection class I Protection level IP24 Electromagnetic compatibility DIN EN 61326 A1 Water supply Destilled or demineralized Water acc to annex C EN 13060 2004 Averaged feed water consumption per cycle ca 0 5 l 2 l Storing conditions Temperature 5 40 C Humidi...

Страница 9: ...t III Module B C1 Power supply Voltage 3N 400V 5 50 Hz 16A Working power 6 kW Averaged power consumption per cycle 6 5 kWh Protection class I Protection level IP24 Electromagnetic compatibility DIN EN 61326 A1 Water supply Distilled or demineralised Water acc to annex C EN 13060 2004 Averaged feed water consumption per cycle ca 0 8 l 2 5 l Storing conditions Temperature 5 40 C Humidity max 85 Heat...

Страница 10: ...ving by sterilizing cooling equalizing finish ventilating gravitation process heating Programs with fast cooling Liquids RM 0 50 100 150 200 250 Time Pressure kPa 0 20 40 60 80 100 120 140 Temp C Athmospheric pressure kPa ventilating equalizing cooling gravitation process finish feed water level check air removing by heating cooling aktiv passiv sterilizing Program Liquids RM in Laboklav ECO with ...

Страница 11: ...mber wall is cooled by air Venting The program is using for the venting function the integrated venting air filter automatically The function works until reaching the programmed air pressure Equalizing For additional safety the device is waiting a few seconds after reaching the program finish to make sure that no measurement mistake has set the program to finish Equalizing time depends on program ...

Страница 12: ... No contact of the heaters with feed water safes live cycle time Protection of the drain on house installation side by using normal tape water mixing to the steam outlet That process works temperature controlled by PT 100 in the outlet installation The Temperature is programmable for opening and closing the mixing valve Enforced cooling option available option M cooling with air Timer controlled p...

Страница 13: ... value at actual cursor position close door 6 Cursor button right moves cursor right 7 Enter button enters the input data or entry in a menu 8 Escape button for leaving a menu position after or before changing is valid 9 Program button for entering the program menu to change the program by using up and down button the program is chosen and activated by pressing the enter button 10 Start button sta...

Страница 14: ...dequate clothes to be protected against burning like temperature stable isolating gloves etc Check the water before putting baskets inside the chamber Fill demineralized feed water up to the height of the bottom sheet inside the chamber manually Preparation of sterilization goods Solid sterilization goods should be cleaned before sterilizing Cleaning is a basic part of sterilization process It red...

Страница 15: ...grams the thermo locking system is activated and protects the user against burning For the use of this mechanism correctly it is necessary to put the reference sensor into a reference flask that is equal in volume and form and filled with the same volume like the biggest single liquid volume inside your chamber Attention please while using hot liquids While contact with hot liquids with temperatur...

Страница 16: ...this The program is going to the next possible program phase without coming in danger for the user or sterilization goods Breaking a program is a special situation for the device The device goes automatically in a standard program phase For liquids does it mean that the unit is switching off the fast cooling function and waits until reaching the removal temperature and removal pressure So that can...

Страница 17: ...st the program with empty chamber and full loaded chamber to be sure the program parameters do not make problems in normal cycle run If there are doubts you should make a full validation of the program Cycle progress The program cycle is running fully automatically The display shows the actual program cycle and gives information what is the actual situation in the running program phase The success...

Страница 18: ...onfigured reference sensor are displayed Date and time are always indicated for documentation issues The program P1 was started but lid was not completely closed So the device generates an error message The error message can be cleared by pressing the Escape button if the cause of the error was cleared The program was started the program has initialized the parameters regular and is now in deaerat...

Страница 19: ...utton Then the lid can be opened and the sterilization good can be removed from chamber In case of a not successful or broken program the sterilizer shows the following message Before regular end of cycle a break was initialized maybe by hand or automatically In case of automatic break an error message is additionally shown Cancellation of program cycle Each program is able to be broken manually b...

Страница 20: ...gram has finished and you quit the program the autoclave is automatically opening the lid Do not get frightened if there is sometimes a loud noise bang during lid opening process This noise is normal Draining the air and steam To remove the air and steam the unit is equipped with an outlet to connect to a drainage system If the unit is connected with cold water supply the program is automatically ...

Страница 21: ...rchive data Protection of the data against changing In following picture you can see the main window of the DOKUMENTATOR software The printer CBM 910II CITIZEN is a dot matrix printer It is available as built in version An interface for external connection is not available With the printer it is possible to printout all relevant process data while the process is running To a later time it is possi...

Страница 22: ...r0040 Disconnected steam generator thermal switch message Disconnected steam gen thermal switch THSZ01 description error occurs if switch THSZ01 is disconnected Er0041 Disconnected chamber thermal switch message Disconnectedchamber thermal switch THZ11 description error occurs if switch THZ11 is opened and switch THSZ01 is closed Er0042 Disconnected safety line message Disconnected safety line des...

Страница 23: ... description error occurs if readout from TIC22 sensor is more than admissible maximum Service check measuring channels CH2 Tref Er0101 Door of sterilizer opened at cycle start message Door of sterilizer is open GS01 description error occurs if controller detects that GS01 switch is not closed on cycle startup see locking symbol in the display Er0106 Exceeded maximum available cycles number for fi...

Страница 24: ... operator without problems Special services can be done by specially trained service stuff only All inspection activities acc to pressure vessel regulations local regulations for pressure vessels and electrical installations need special trained service stuff We recommend to order one regular safety inspection per year and to connect this with a regular maintenance for the vessel pressure parts an...

Страница 25: ...ERVICE Check program parameters X X Check for lid door adjustment X SERVICE Electrical test BGVA 2 4 X SERVICE Attention please Opening the unit is allowed for authorized and trained service stuff only Cleaning Before starting with cleaning the device please disconnect the unit from power supply completely Cleaning should be done if the unit was cooling down only Danger if the chamber is hot Clean...

Страница 26: ...afety valve s needs to be checked once per year This should be done by specially trained service stuff Other safety checks are necessary so we recommend making one safety check together with the yearly maintenance check by a trained service engineer While testing the function of the safety valve steam is leaving the safety valve Attention please While steam is leaving the safety valve the valve is...

Страница 27: ...Pa absolute pressure 2 45 bar relative pressure an alarm is generated including error message and the heating function is switched of and the unit breaks the program With 2 8 bar relative pressure the safety valve is opening and chamber pressure is reduced mechanically The steam is blowing into the housing contact with the steam is not very dangerous because it is saturated after blowing out Atten...

Страница 28: ... EN 285 Steam sterilizers app B EN 13060 small size steam sterilizers App C Feed water Condensate Residual dry matter 10 mg l 1 0 mg kg Silica oxide SiO2 1 mg l 0 1 mg kg Iron 0 2 mg l 0 1 mg kg Cadmium 0 005 mg l 0 005 mg kg Lead 0 05 mg l 0 05 mg kg Other heavy metals except for iron cadmium lead 0 1 mg l 0 1 mg kg Chlorines 2 mg l 0 1 mg kg Phosphates 0 5 mg l 0 1 mg kg Conductivity at 20 C 15 ...

Страница 29: ...ce company or is able to solve your problem directly Manufacturer Germany SHP Steriltechnik AG Schloss Detzel 1 39345 Detzel Schloss Fax 49 39058 97 62 22 08 00h 17 00h Mon Fr Tel 49 39058 97 62 0 08 00h 17 00h Mon Fr Mobil 49 177 6269880 24 h Email info shp steriltechnik de 08 00h 17 00h Mon Fr You can also contact the local importer distributor as per your invoice or the mentioned on the sticker...