LABOKLAV Eco 80 / 135
User Manual
26
- Cleaning the chamber inside – For cleaning the chamber use a wet and soft towel from
cotton material. Special cleaning material or chemicals are not necessary. Do not use aggressive or
organic chemical for cleaning! Chemicals can damage the sealing or sensors!
The rest of chemicals or cleaning materials will be brought
forward to the sterilization goods of next sterilization cycle!
Do not use aggressive or organic chemicals for cleaning!
- Cleaning the accessories – Clean the baskets etc. with wet towel or under running water.
- Cleaning the housing – The housing needs to be cleaned by wet towel or light oil. Special
cleaning chemicals like used in house hold can be used.
Check of the safety valve
The safety valve(s) needs to be checked once per year. This should be done by specially trained
service stuff. Other safety checks are necessary so we recommend making one safety check
together with the yearly maintenance check by a trained service engineer. While testing the function
of the safety valve steam is leaving the safety valve.
Attention please! While steam is leaving the safety valve the
valve is getting very hot in a very short time! Make sure that
your human skin is not coming in contact with the steam!
Contact with the steam can burn your skin!
If the valve is not closing completely after testing it needs to
be changed! If there are doubts about the regular functionality
of the tested valve it needs to be changed!
Changing the venting air filter
Wear and tear of the venting air filter depends on the number of cycles and the quality of the
environment. We recommend changing the filter after 100 cycles or once per month.