29
(12) Move the Pen down to close the Door.
Move the pen down with the pen-up lever. Close the door and check that it is locked.
The pen-up lever is located at the inside right end of the case. Push it down using a finger.
(13) Press the “FEED” key to check Paper Feed.
Press the “FEED” key on the display keyboard to feed the chart paper.
Fig. 5.1-13 Pressing the “FEED” key
(14) Press the “REC” key to restart operation.
The chart paper is fed
while pressing the key.
Releasing the key stops
paper feed. The paper
begins a slow feed when
the key is pressed.
Prior to the operation of
the instrument, be sure to
confirm that the chart
paper is being fed.
L
M
Caution
Be careful when opening the door more than 90 degrees. Also make sure that the display cable
does not get caught.