104
(1) Calibration of Voltage
Conduct calibration for voltage, current or thermocouple input. For thermocouple input, calibrate
reference junction compensation together.
Conduct calibration according to the instructions in Section 7.2.12 Data Calibration (Calibration of the
voltage). (See p.91)
(2) Calibration of RTD (Resistance Temperature Detector)
Conduct calibration according to the instructions in 7.2.12 Data Calibration (Calibration of the RTD).
(See p.91)
Precision Dial
Resistor
HR-701 (702) Recorder
Input Terminals
Wiring
A
B
B
Wiring
Precision Voltage
(Current)
Generator
+
-
Digital Voltmeter
HR-701 (702) Recorder
Input Terminals
+
-
+
-
[Note]
The dispersion of resistance for each wire must be 6 m
or less.
[Note]
Calibrate at every channel.
When a shunt resistor is attached to the input terminal block in the case of current input, detach it
prior to calibration.