99
8.4 Outputting the Unit and Decimal Point Position Data
8.4.1 Specifying the Unit and Decimal Point Position Data Output
TS2(CR) (LF)
Once this command is received, the recorder transfers the unit and decimal point position data to the
transmission buffer, when the (ESC) T command is received.
8.4.2 Updating the Data
(ESC)T(CR) (LF)
Once this command is received, the recorder transfers the updated unit and decimal point position
data to the transmission buffer.
8.4.3 Outputting the Data
LF, S CH, E CH (CR) (LF)
S CH: Output starting channel
01 to 02
E CH: Output ending channel
01 to 02
(e.g.) LF, 01, 02(CR) (LF) Unit and decimal point place data output for Ch1 and Ch2
When the number of channels is 1, specify (S_CH) = (E_CH).
8.4.4 Data Format
(DS
1
)(DS
2
)(CHNo.) (UNIT
1
to
6
)(DP) (CR) (LF)
(1) DS
1
: Data information 1 (1 byte)
N : Normal
D : Difference computation data
S : Sum computation data
M : Average computation data
R : Square root computation data
C : Decade computation data
O : Overrange (Data =
99999)
S : Skip (All data: Space)
(2) DS
2
: Data information 2 (1 byte)
E: Final data
(Space): Other data
(3) CH No.: Channel number (2 bytes);
01 to 02
(4) UNIT
1 to 6
: UNIT
The set unit is returned by codes.
(When sending the 7-bit data, 1 bit information is omitted.)
(5) DP: Decimal point information (1 byte);
0 to 4
[Note]
Depending on the range setting, a decimal point position may differ from that of the display screen.