CAUTION
●The
mark on the plate affixed to the instrument on the terminal nameplate affixed to the case of your
instrument, the
mark is printed. This is to warn you of the risk of electrical shock which may result if the
charger is touched while it is energized.
●The external power circuit connected to the power terminal of this instrument must have a means of
turning off the power, such as a switch or breaker. Install the switch or breaker adjacent to the instrument in a
position which allows it to be operated with ease, and with an indication that it is a means of turning off the
power. Use a switch or breaker, which meets the requirements of IEC127.
●Fuse
Since the instrument does not have a built-in fuse, do not forget to install a fuse in the power circuit to be
connected to the power terminal. The fuse should be positioned between the switch or breaker and the instrument
and should be attached to the L side of the power terminal.
Fuse Rating: 250V AC 0.5A medium lagged or lagged type.
Use a fuse which meets the requirements of IEC127.
●Load voltage/current to be connected to the output terminal and the alarm terminal should be within the
rated range. Otherwise, the temperature will rise and shorten the life of the product and/or result in problems with
the product.
●Voltage/current that differs from input specification should not be connected to the input terminal. It may
shorten the life of the product and/or result in problems with the product.
●Input, DI, AI, CT, and FB terminals are not insulated. When use a grounding-type sensor, please do not ground
with those terminals.
●A signal wire's common mode voltage to ground (signal wires other than contact output including power
supply and event) should be less than 30V rms, 42.4V peak, and 60 VDC.
●All the wires for the interior distribution, except for communication and contact output (including power supply
and event), should be less than 30m in length. When the wire's length is 30m or more, or in the case of
outdoor wiring, the suitable measure against a lightning surge is required.
●EMC standard (IEC61326) classifies MAP6 into Class A apparatus. Electromagnetic interference may occur
when MAP6 is used at a business district or in the home. Please use after taking sufficient measures.
●There is a ventilation hole for radiation. Please make sure that the metal won't enter from this hole. It
may be occur a breakdown and fire.
●Do not block the draft hole or allow dust or the like to stick to it. A rise in temperature or insulation
failure may result in a reduction of the life of the product and/or problems with it or may cause a fire.
●It should be noted that repeated tolerance tests against voltage, noise, surge, etc. may lead to deterioration
of the instrument.
●Users are prohibited from remodeling the product or abnormal use thereof.