Firmware Recovery
30
SHARP
Document Solution Centre
6.
Using the [UP] and [DOWN] keys, navigate and select
the required firmware SFU file .
Use the [OK] key to browse through to the sub folder(s) or
[BACK] key to return to the parent folder(s).
7.
Press [OK]. – It will take approx. 1-2 minutes to read the
SFU file.
8.
After reading is completed, the firmware update (writing)
will commence.
Note:
The abbreviated name/location of the firmware that is being
updated will appear in the upper right corner.
9.
If the firmware upgrade completes successfully,
RESULT OK will be displayed.
10.
Power off the MFP. Pull out the MFP PWB and
Turn OFF MFP PWB UPDATE DIP SW.
11.
Insert and secure the MFP PWB and refit the Right Rear
Cabinet.
Firm Update
> 0100h200.sfu
Firm Update
Reading Data
Firm Update IcuM
Writing Data
Firm Update IcuM
Result : OK