5-5. F-Function Detailed information
■ General Function Setting (“●” Factory default set value)
Weight-Back up selection
F02
0
Normal Mode
●
1
Weight Back up Mode
Motion Band Range setting
F03
06
00
∫
50
This is set “Steady” acceptable range of weighing part.
If there is vibration on weighing part, you can set this function and reduce the vibration
effect on weighing process. (0 : Weak vibration ~ 50 : Strong Vibration)
Zero Tracking Compensation Range setting
F04
02
00
∫
09
Due to external causes(Temperature, wind, and dust), there are small weight difference,
indicator will ignore the weight difference and display Zero.
For this compensation function, indicator will estimate the weight difference is over the
set range during fixed time period.
If there is large weight difference over set range within fixed time period, the “Zero” is
breaking and will find new zero point.
Auto Zero Range setting
F05
00
00
∫
99
Within the “Auto Zero” range, weighing part is steady, indicator will display current
weight as “Zero” If the weighing part is not “Steady”, indicator will display current
weight. (Auto Zero Range : ± Set value + weight unit)
Digital Filter setting
F06
13
AB
A : Frequency Filter setting value (0~1)
(0 : about 200Hz/sec, 1 : about 500Hz/sec)
B : Buffer Filter setting value (1~9)
If “B” set value is fixed, “A” set value is
large, the indicator will response more
sensitive.
Zero /Tare key Operation mode selection
F07
●
0
Activate when “Steady” condition, only
1
Always activated
Zero key Operation Range selection
F08
0
Activated within 2% of Max. Capacity
1
Activated within 5% of Max. Capacity
●
2
Activated within 10% of Max. Capacity
3
Activated within 20% of Max. Capacity
4
Activated within 30% of Max. Capacity
5
Activated within 50% of Max. Capacity
6
Activated within 100% of Max. Capacity
Caution : If you set over 20% , there may be "CELL-Err" or displaying wrong value.