DIGITAL WEIGHING INDICATOR
SI 4000
45
7-3. Digital Weighing Indicator
Error
No.
Display
Cause
Treatment
No.1
“CELL-Er”
or
“--OL--”
1. Load cell Error
2. Load cell cable Error
3.Load cell connection Error
4. A/D Board Error
5.It displays under 5000 or
Over520000.
1. Under “TEST” mode 1, check analogue
value. If you can not get any analogue value
or there is no change although adding load,
please check load cell, load cell cable,
connection conditions first.
2. Replace another load cell, and check the
indicator condition. If you have same
problem, please replace new indicator and
check A/D board error.
No.2
“Un-Pass”
1. Power is ON, when some
materials are on weighing part.
※
Under “Normal Mode”, if there
are more than 20% loading of
Max. capacity, “
Un-Pass
”
display will be appeared and
indicator will stay until
removing the load.
※
Setting Back-up mode it can
memory empty value, and it
becomes set value without
displaying” Un-pass”)
1. If you set “Normal Mode”, please check
weighing part empty or not before turn on
the power. If there are some materials in/on
weighing part, please remove those
materials and turn on the power.
2. Please try to set F02-01(Back-up) mode so
that the indicator can remember first
empty value.
No.3
“FN-SET”
1. When “FN-Memory” is defected
2. When the “FN-Memory” is
empty.
1. Please contact the distributor or Head Office.
No.4
“P-Err”
Under Parallel Printer is connected
and installed.
1. Parallel printer interface is
defected or disconnected.
1. Please check connection of the print cable.
2. Please check the trouble of print.
※
If you only install “Parallel Print” option
card, you can check to do.
※
Under “CELL-Er”, Relay will not be Output, and Analogue Output(4~20mA/0~10V), either.