DIGITAL WEIGHING INDICATOR
SI 4000
17
6. Define the optimal position or decimal point
Whenever you press
key, the location of decimal point
will be changed.
p 30.00
7. Press
key to save Digit / Decimal point and
move to next step.
Ex) Load cell CAPA : 30kg, division : 0.01
Input 30.00
p 30.00
8. Define the optimal Digit/Division value of weighing measurement.
Whenever you press
key, the Digit/Division value will
be changed in order of “1
2
5
10
20
50” .
Ex) Load cell CAPA : 30kg, division : 0.01
Input division “1”
d 1
9. press
key to save the Digit/Division value and
move to next step.
d 1
※
Caution : (Division value /Max capacity value)
cannot over
1/20,000
.
If the division is so small compare with max capacity, Error message
“ Err 01 “
will be displayed
and move back to
“CAPA”
mode again.
10. Under this step, measure the “DEAD Weight of Weighing Scale
When you press
key, the indicator starts the dead process
to find “Zero” span.
dead
11. Indicator will search “DEAD weight” during 5seconds.
After find optimal “Zero” span , step is automatically
Moves to next.
Cal-1 0