DIGITAL WEIGHING INDICATOR
SI 4000
15
12. Input the weight of your “Test weight”. And press
key.
Ex) Load Cell CAPA : 20kg, division 0.001
Use test weight which is at least 10% of
max CAPA(20kg) = minimum 2kg of test weight is needed.
Input test weight 2.000 to indicator.
l 2.000
13. When “UP” is displayed, load your test weight on the scale (weigh
bridge) Ex) Load Cell CAPA : 20kg, division 0.001
Up
14. And press
key.
Do not remove the test weight from weigh bridge.
Up
15. Indicator will calculate span value during 5sec.
Cal-2 0
16. After finish calculation, span value will be displayed.
Please remove the test weight from weigh bridge.
0.629238
※
Caution :
The
“Test Weight’s value”
must be at least
10%
max. capacity of weighing scale.
“at least 10%” means to guarantee precise weighing process you have to make standard with at least
10% of the max capacity weight.
We programmed the calibration will not be done, when you load less than 10% of the max capacity.
17. Press
key to save all calibration process.
After then it resets automatically.
Now, fasten the Calibration Bolt on the front panel.
end