SERVOTOUGH SpectraScan User Manual
02400001A/1
67
11 TROUBLESHOOTING
See Table 6 - Diagnostic LED Indications for a general description of each indicator status. Below
are some common problems/symptoms and possible causes.
Problem
Possible cause(s)
Power LED not lit
No or insufficient power. Check Fuse
Warm-up light still lit after 2 hours of warm up
period
Environment is too cold to allow stabilization at
the operating temperature
Negative concentration reading
Zeroing on sample gases or insufficient purging
prior to zeroing
Excessive drift
Analyzer has not fully warmed up or sensor
head temperature is higher than its intended
operating temperature (default 60° C)
Unable to communicate with the analyzer
Incorrect IP address
Low light indicator (<50%)
Dirty windows or light source has degraded
No measurement response or unchanging
measurement values
Unconnected or improperly connected sample
line, blocked flow
Slow measurement response
Low flow rate due to blockage
Table 7 - Troubleshooting
Содержание Servotough SpectraScan 2400
Страница 7: ...SERVOTOUGH SpectraScan User Manual 02400001A 1 v This page intentionally left blank...
Страница 12: ...SERVOTOUGH SpectraScan User Manual 02400001A 1 10...
Страница 13: ...SERVOTOUGH SpectraScan User Manual 02400001A 1 11 This page intentionally left blank...
Страница 30: ...SERVOTOUGH SpectraScan User Manual 02400001A 1 28 This page intentionally left blank...
Страница 33: ...SERVOTOUGH SpectraScan User Manual 02400001A 1 31 Figure 4 Analyzer with Cover Open...
Страница 58: ...SERVOTOUGH SpectraScan User Manual 02400001A 1 56 This page intentionally left blank...
Страница 60: ...SERVOTOUGH SpectraScan User Manual 02400001A 1 58 This page intentionally left blank...
Страница 68: ...SERVOTOUGH SpectraScan User Manual 02400001A 1 66 This page intentionally left blank...
Страница 70: ...SERVOTOUGH SpectraScan User Manual 02400001A 1 68 This page intentionally left blank...