SERVOTOUGH SpectraScan User Manual
02400001A/1
26
5.0.3
Combustion Hazard
The Servomex SpectraScan Gas Analyzer is rated for certain combustible environments. Ensure
that the installation location of the analyzer aligns with the certifications listed on the product’s
labelling.
5.0.4
Power Requirements
The system is designed for 24 VDC operation. An optional AC Power Module can be installed at the
factory for use with 100-240 VAC. The main power board is fused for use with a 24 VDC supply.
Refer to Table 13 - Electrical Specifications on page 72 for more detailed information.
5.0.5
Fittings Requirements
The Servomex SpectraScan Gas Analyzer comes standard with ¼” male Swagelok
tube fittings on
the sample inlet and exhaust ports. Only ¼” female Swagelok
tube fittings should be used on the
incoming and outgoing sample lines, to mate with the Servomex SpectraScan sample inlet and
exhaust ports. Use of stainless steel tubing for the incoming and outgoing sample lines is
recommended. Instructions for installing the female Swagelok tube fittings onto the incoming and
outgoing sample lines are provided in Section 7 SAMPLE GAS CONNECTIONS on page 49.
Other fitting sizes and styles are available from Servomex to be installed on the inlet and exhaust
ports as a special order. The underlying connector is a ¼” NPT.
CAUTION
Equipment Damage: Provide adequate support for the tubing. The weight of the tubing must not
exert torque on the gas cell as it may cause the fittings to fail and leak.
All of the fittings requirements for the Servomex SpectraScan Gas Analyzer are summarized below.
Connector on the Analyzer
Required Fittings to
Mate with Analyzer
Additional Comments
Sample Inlet And Exhaust
Ports:
¼” male Swagelok
¼” female Swagelok
10 micron particulate filter on the
inlet side is recommended.
Electrical Conduit:
1” female
1” male conduit
Must maintain Ingress Protection
level of IP66, Type 4.
Purge Ports (use is optional):
¼” female NPTF
¼” male NPTF
Either side can be used as inlet.
Table 3 - Standard Fittings Used on the Servomex SpectraScan Gas Analyzer
5.1
Mounting Procedure
1. After choosing the appropriate space and structure for the Servomex SpectraScan Gas
Analyzer by satisfying all of the
Environmental Requirements
on page 23, prepare the unit
for mounting by clearing the mounting space, readying the analyzer (consider its weight and
Содержание Servotough SpectraScan 2400
Страница 7: ...SERVOTOUGH SpectraScan User Manual 02400001A 1 v This page intentionally left blank...
Страница 12: ...SERVOTOUGH SpectraScan User Manual 02400001A 1 10...
Страница 13: ...SERVOTOUGH SpectraScan User Manual 02400001A 1 11 This page intentionally left blank...
Страница 30: ...SERVOTOUGH SpectraScan User Manual 02400001A 1 28 This page intentionally left blank...
Страница 33: ...SERVOTOUGH SpectraScan User Manual 02400001A 1 31 Figure 4 Analyzer with Cover Open...
Страница 58: ...SERVOTOUGH SpectraScan User Manual 02400001A 1 56 This page intentionally left blank...
Страница 60: ...SERVOTOUGH SpectraScan User Manual 02400001A 1 58 This page intentionally left blank...
Страница 68: ...SERVOTOUGH SpectraScan User Manual 02400001A 1 66 This page intentionally left blank...
Страница 70: ...SERVOTOUGH SpectraScan User Manual 02400001A 1 68 This page intentionally left blank...