Figure 13 - Sample session in which a static IP address is set using PuTTY to
communicate with the MPU.
6.4.4
RS-422 connection
The RS-422 link uses a 4-wire connection, 2 wires for receive and 2 wires for transmit. The RS-422
port supports only a point-to-point link. Do not bridge the receive and transmit pairs – the
instrument cannot be used in multi-drop bus configurations.
Connecting the instrument's RS-422 port to an external RS-422 device requires a 5 conductor
cable. The cable should include two twisted pairs for the receive and transmit data, and a single
wire for data ground. The configuration is shown in Drawing 1. Connection to the system board is
made using the connector labelled “RS-485” at the corner of the board.
Содержание Servotough SpectraScan 2400
Страница 7: ...SERVOTOUGH SpectraScan User Manual 02400001A 1 v This page intentionally left blank...
Страница 12: ...SERVOTOUGH SpectraScan User Manual 02400001A 1 10...
Страница 13: ...SERVOTOUGH SpectraScan User Manual 02400001A 1 11 This page intentionally left blank...
Страница 30: ...SERVOTOUGH SpectraScan User Manual 02400001A 1 28 This page intentionally left blank...
Страница 33: ...SERVOTOUGH SpectraScan User Manual 02400001A 1 31 Figure 4 Analyzer with Cover Open...
Страница 58: ...SERVOTOUGH SpectraScan User Manual 02400001A 1 56 This page intentionally left blank...
Страница 60: ...SERVOTOUGH SpectraScan User Manual 02400001A 1 58 This page intentionally left blank...
Страница 68: ...SERVOTOUGH SpectraScan User Manual 02400001A 1 66 This page intentionally left blank...
Страница 70: ...SERVOTOUGH SpectraScan User Manual 02400001A 1 68 This page intentionally left blank...