SERVOTOUGH SpectraScan User Manual
02400001A/1
44
2.
Connect a Windows PC to the MPU debug connector using the SpectraScan RS-232 cable
(and an RS-232 to USB converter, if required).
3.
On your PC, determine which COM port is associated with the serial connection. Do this by
opening the Windows device manager and looking at the COM port configuration.
4.
On your PC, run your terminal program and configure it to use the COM port you found in the
device manager. Also configure the terminal program with the following settings:
Baud rate = 115200
Data bits = 8
Stop bits = 1
Parity = None
Flow control = None. (No software nor hardware flow control.)
5.
In the terminal program, start a terminal session to the MPU. Hit <return> a few times and you
will get a command prompt from the MPU. The command prompt is “#”.
6.
At the command prompt, type the command “ip-config” followed by <return>. When the
program starts it will tell you the current IP configuration of the unit. Also, if the unit is already
configured for static IP address, it will give you the option to leave the program.
7.
Next, the program will give you three options to modify the IP address settings. Choose option
2, “static IP”.
8.
The program will
9.
When prompted, type in your desired IP address is DD.DD.DD.DD format. (Here, DD signifies
the digits comprising each of the four decimal numbers making up the IP address.)
10. Next, you will be prompted for the desired netmask. Enter it using the DD.DD.DD.DD format.
Figure 12 - Location of debug RS-232 connector on MPU
Содержание Servotough SpectraScan 2400
Страница 7: ...SERVOTOUGH SpectraScan User Manual 02400001A 1 v This page intentionally left blank...
Страница 12: ...SERVOTOUGH SpectraScan User Manual 02400001A 1 10...
Страница 13: ...SERVOTOUGH SpectraScan User Manual 02400001A 1 11 This page intentionally left blank...
Страница 30: ...SERVOTOUGH SpectraScan User Manual 02400001A 1 28 This page intentionally left blank...
Страница 33: ...SERVOTOUGH SpectraScan User Manual 02400001A 1 31 Figure 4 Analyzer with Cover Open...
Страница 58: ...SERVOTOUGH SpectraScan User Manual 02400001A 1 56 This page intentionally left blank...
Страница 60: ...SERVOTOUGH SpectraScan User Manual 02400001A 1 58 This page intentionally left blank...
Страница 68: ...SERVOTOUGH SpectraScan User Manual 02400001A 1 66 This page intentionally left blank...
Страница 70: ...SERVOTOUGH SpectraScan User Manual 02400001A 1 68 This page intentionally left blank...